Rosemount Process Radar
5-20
Failure Rate Data
The FMEDA report includes failure rates and common cause Beta factor
estimates. The full report is accessible at
www.emersonprocess.com/rosemount/safety/PriorUse.htm.
Useful Lifetime
The established failure rates of electrical components apply within the useful
lifetime, which should be based on experience. According to IEC 61508-2,
7.4.7.4, note 3, the useful lifetime often lies within a range of 8 to 12 years for
transmitters.