
Reference Manual
00809-0600-4026, Rev BA
December 2011
Rosemount Process Radar
4-4
Probe End Projection
Probe End Projection allows you to measure product level when the surface
pulse is too weak to be detected, such as long measuring ranges, solids, and
boiling hydrocarbons. The calibration should be completed first with an empty
tank during cold commissioning, and then a second time with a filled tank, but
do not overwrite the empty tank calibration.
1. Make sure the surface echo is above the threshold
2. In Rosemount Radar Master click Guided Probe End Projection and
follow instructions
For more information, see the Rosemount 5300 Series Reference Manual
(Document No. 00809-0100-4530) section “Probe End Projection”.
NOTE
Probe End Projection is not compatible with Signal Quality Metrics, Dynamic
Vapor Compensation, Interface Measurement, and High DC applications.
Verification Procedure
Step 1. Verify the
Correctness of Level
Reading
At normal operating conditions, compare the transmitter level, or interface
reading, with an independent measurement. Unfortunately, it often happens
that two independent measurements will not match up perfectly, but check the
sanity and the acceptable deviation.
The independent measurement can be done using a number of different
complementary devices, ranging from sight-glasses and hand-dipping to
redundant differential-pressure and displacer.
Figure 4-2. The verification of
measurement data using a
sight-glass
Kommentare zu diesen Handbüchern