Reference Manual
00809-0600-4026, Rev BA
December 2011
Rosemount Process Radar
3-2
INSTALLATION &
CONFIGURATION
Step 1. Complete the QIG Follow the instructions provided in Section 2: 5300 Quick Installation Guide.
Step 2. Complete the
Basic Configuration
Basic Configuration Parameters
HART/FF Address
Device Tag
Measurement Units
Tank Geometry
Figure 3-1. Tank Geometry
Probe Type
Hold Off/Upper Null Zone
This parameter should only be changed if there are disturbing objects
close to the antenna, e.g. nozzle disturbances. No valid measurements
are possible above the Hold Off Distance.
Measurement Mode
Rapid Level Changes
Dielectric Constant
Volume Configuration
For volume calculations, you can choose one of the standard tank
shapes or the strapping option. Choose None if volume calculation is
not used.
Analog Output (HART)
Tank Height
Hold Off/
Upper Null Zone
Product Level
Upper Reference Point
Lower Reference Point
Interface
Level
Probe
Length