Reference Manual
00809-0600-4026, Rev BA
December 2011
Rosemount Process Radar
4-6
Step 3. Echo-Curve
Verification at Operating
Conditions
At normal operating conditions, download and review the echo-curve
according to the following steps:
1. Check Reference Pulse amplitude and position
Compare the result against the previous plots, taken during cold start-up.
2. Review peak amplitudes and threshold settings
Verify that the surface and/or interface peaks are visible and that the
thresholds have been set as described in threshold settings. See table
below for rough guidelines of amplitudes. For information on threshold
settings, see “Threshold settings“ on page 4-3.
3. Store the echo-curve
Permanently store the echo-curve for future use, with re-occurring
verification procedures.
Figure 4-5. Peak amplitudes and
threshold settings
Table 4-1. Typical peak
amplitudes for Rosemount 5300
Series with single lead probe in
4 in. chambers
Common Problems and Recommended Actions
Surface pulse not detected
UNZ
Surface
peak
Interface
peak
Interface
threshold
Surface
threshold
Reference
peak
Distance
Amplitude
Peak
Approximate signal strength, ideal
conditions for single lead probe in 4 in.
(100 mm) chambers
Reference peak
~10,000 mV
(1)
(1) This value does not apply and may be considerably lower when the probe is completely submerged
in product.
Surface peak, 5301 with oil (DC=2) ~2,000 mV
Surface peak, 5301 with water (DC=80) at
3 ft (1 m) distance
~10,000 mV
Interface peak, 5302 with oil and water ~8,000 mV