Seitenansicht 95
Instruction Manual
Appendix BX Rev. 1.2
January 1997
B-8 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
MAINTENANCE AND SERVICE
Consult Safety Data Sheet 1M03243
before performing any work on the
CENELEC approved HPS 3000.
Install all protective equipment covers
and safety ground leads after equip-
ment repair or service. Failure to in-
stall covers and ground leads could
result in serious injury or death.
B-7 OVERVIEW
This section describes service and routine
maintenance of the HPS 3000 Heater Power
Supply Field Module. Replacement parts re-
ferred to are available from Rosemount. Refer
to Replacement Parts of this appendix for part
numbers and ordering information.
B-8 FUSE REPLACEMENT
The heater power supply's motherboard (25,
Figure B-6) contains four identical 5 amp fuses
(24) (5 amp anti-surge, Type T to IEC127,
ROSEMOUNT Part No. IL01293H02). To check
or replace a fuse, simply unscrew the fuse
holder cap (23) with a flat head screwdriver and
remove fuse. After checking or replacing a fuse,
reinstall fuse holder cap.
B-9 TRANSFORMER REPLACEMENT
Consult Safety Data Sheet 1M03243
before performing any work on the
CENELEC approved HPS 3000.
a. Follow the power down procedure outlined
in safety data sheet 1M03243 and the offi-
cial "Codes of Practice" for your country of
installation.
b. Loosen setscrew (13, Figure B-6) retaining
HPS cover (14). Remove HPS cover.
c. Remove hex nut (2), flat washer (3), and
hex head bolt (4) that retain transformer to
mounting bracket.
d. Disconnect the 6-pin transformer wiring
harness connector from J7 on the HPS
motherboard (25). Disconnect the 2-pin
wiring harness connector from the jumper
that was selected to configure the HPS for
the proper voltage.
NOTE
The transformer connector is keyed
making connection possible in only
one direction.
e. Remove old toroidal transformer (11). Place
new transformer in position and reconnect
wiring harness connectors as noted in
step d.
f. Place insulating disc (10) (one on either
side of transformer) and metal disc (12) on
transformer.
g. Tighten hex nut (2) and hex head bolt (4)
only enough to firmly hold transformer in
place.
h. Reinstall HPS cover (14) and secure with
setscrew (13).
B-10 MOTHERBOARD REPLACEMENT.
Consult Safety Data Sheet 1M03243
before performing any work on the
CENELEC approved HPS 3000.
a. Follow the power down procedure outlined
in safety data sheet 1M03243 and the offi-
cial "Codes of Practice" for your country of
installation.