Instruction Manual
106-300NFX Rev. 4.2
January 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Description 1-3
World Class 3000
CENELEC approved probes are available in
three length options, giving the user the
flexibility to use an in situ penetration ap-
propriate to the size of the stack or duct.
The options on length are 457 mm (18 in.),
0.91 m (3 ft), and 1.83 m (6 ft). The probe is
certified EExd IIB T1 [370°C (698°F)] to
CENELEC standards EN50014 and
EN50018.
The IFT contains electronics that control
probe temperature (in conjunction with the
optional HPS) and supply power, and pro-
vide isolated outputs that are proportional to
the measured oxygen concentration. The
oxygen sensing cell is maintained at a con-
stant temperature by modulating the duty
cycle of the probe heater. The IFT accepts
millivolt signals generated by the sensing
cell and produces outputs to be used by
remotely connected devices. The IFT output
is isolated and selectable to provide
linearized voltage or current.
The heater power supply CENELEC
approved (HPS) can provide an interface
between the IFT and the probe. The HPS
contains a transformer for supplying proper
voltage to the probe heater. The unit is
certified EExd IIC T6 to CENELEC stan-
dards EN50014 and EN50018.
Systems with multiprobe and multiple IFT
applications may employ an optional MPS
3000 Multiprobe Test Gas Sequencer. The
MPS 3000 provides automatic test gas
sequencing for up to four probes and IFTs
to accommodate automatic calibration. The
MPS 3000 must be installed in a non-
hazardous, explosive-free environment.
d. System Features
1. Unique and patented electronic cell
protection action that automatically
protects sensor cell when the analyzer
detects reducing atmospheres.
2. Output voltage and sensitivity increase
as the oxygen concentration de-
creases.
3. User friendly, menu driven operator
interface with context-sensitive on-line
help.
4. Field replaceable cell.
5. Analyzer constructed of rugged 316
LSS for all wetted parts.
6. The intelligent field transmitter (IFT)
can be located up to 45 m (150 ft) from
the probe when used without optional
heater power supply (HPS). When the
system includes the optional HPS, the
HPS can be located up to 45 m (150 ft)
from the probe and the IFT may be lo-
cated up to 364 m (1200 ft) from the
HPS.
7. All electronic modules are adaptable to
120, 220, and 240 line voltages.
8. Five languages may be selected for
use with the IFT. These are:
English
French
German
Italian
Spanish
9. An operator can set up, calibrate, or
troubleshoot the IFT in one of two
ways:
(a) Optional General User Interface
(GUI). The GUI is housed within
the IFT electronics enclosure and
makes use of an LCD display and
keypad.
(b) Optional LED Display Panel (LDP).
The LED display and a limited
function keypad permit calibration
only.