Instruction Manual
Appendix AX Rev. 2.1
February 1998
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices A-13
World Class 3000
l. Reinstall probe and gasket on stack flange.
If there is an abrasive shield in the stack,
make sure dust seal gaskets are in place as
they enter 15° reducing cone.
m. Consult Safety Data Sheet 1M03226 then
turn power on to electronics and monitor
thermocouple output. It should stabilize at
29.3 mV ±0.2 mV. Set reference air flow at
56.6 L/hr (2 scfh). After probe stabilizes,
calibrate probe per Instruction Bulletin ap-
plicable to your electronics package. If new
components have been installed, repeat
calibration after 24 hours of operation.
A-12 CERAMIC DIFFUSION ELEMENT
REPLACEMENT
NOTE
This refers to ceramic diffuser element
only.
a. General
The diffusion element protects the cell from
particles in process gases. It does not nor-
mally need to be replaced because the vee
deflector protects it from particulate erosion.
In severe environments, the filter may be
broken or subject to excessive erosion. Ex-
amine the ceramic diffusion element when-
ever removing the probe for any purpose.
Replace if damaged.
Damage to the ceramic diffusion element
may become apparent during calibration.
Compare probe response with previous re-
sponse. A broken diffusion element will
cause a slower response to calibration gas.
Hex wrenches needed to remove setscrews
and socket head screws in the following
procedure are available as part of a Probe
Disassembly Kit, Table A-3.
Wear heat resistant gloves and cloth-
ing to remove probe from stack.
Normal operating temperatures of
diffuser and vee deflector are
approximately 316° to 427°C (600° to
800°F). They can cause severe burns.
Before carrying out this procedure,
consult Safety Data Sheet 1M03226.
Do not remove cell unless it is certain
that replacement is necessary. Cell
cannot be removed for inspection
without damaging it.
b. Replacement Procedure
1. Follow the power down procedure out-
lined in Safety Data Sheet 1M03226
and the official "Codes of Practice" for
your country of installation. Disconnect
cable conductors and remove cable,
Figure A-10. Shut off and disconnect
reference gas and calibration gas sup-
plies from probe junction box. Wearing
heat resistant gloves and clothing,
carefully remove probe assembly from
stack.
2. Loosen setscrews, Figure A-13, using
hex wrench from Probe Disassembly
Kit, Table A-3 and remove vee deflec-
tor. Inspect setscrews. If damaged, re-
place with stainless setscrews coated
with anti-seize compound.
3. On systems equipped with abrasive
shield, remove dual dust seal gaskets.
AX