Instruction Manual
Appendix AX Rev. 2.1
February 1998
A-12 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
c. Loosen 4 socket head screws from retainer.
Pull off retainer and calibration tube in-
sert, Figure A-12. Discard calibration tube
insert.
d. Press the cell flange against the probe end
flange and twist through 90° in both the
clockwise and counterclockwise directions.
This breaks the bond between the cell and
the platinum pad. Once the bond is broken,
remove the cell.
e. View the platinum pad through the probe
end flange. Reform the platinum pad into a
shape to match the cell electrode by gath-
ering in any loose strands which may have
become untidy. This may be done with a
small flat screwdriver.
f. Remove and discard corrugated seal. Clean
mating faces of probe tube and retainer.
Remove burrs and raised surfaces with
block of wood and crocus cloth.
g. Rub one or two drops of anti-seize com-
pound #3535B53G01 between fingers.
Smear on both sides of corrugated seal.
h. Install new calibration tube insert in retainer,
with the short arm of the tube insert pene-
trating the calibration gas passage in the
outer ring of the retainer. Assemble retainer,
cell and flange assembly, corrugated seal,
PROBE
TUBE
PIN SPANNER
WRENCH
SNUBBER
DIFFUSOR
19780017
Figure A-11. Removal of Snubber Diffuser
and probe tube. Make sure the calibration
tube insert lines up with the calibration gas
passage in each component. Firmly seat
the calibration tube insert in the assembly.
Apply a small amount of anti-seize com-
pound to screw threads and use screws to
secure assembly. Torque to 4 N·m
(55 in-lbs).
i. To install snubber diffuser, apply anti-seize
compound to threads of probe tube, Figure
A-11, and snubber diffuser. Reinstall dif-
fuser on probe tube. Using pin spanner
wrenches, torque to 14 N·m (10 ft-lbs).
j. If using the optional vee deflector and ce-
ramic hub, apply anti-seize compound to
threads of probe end flange, hub, and
setscrews. Reinstall hub on probe tube.
Using pin spanner wrenches, torque to
14 N·m (10 ft-lbs). Reinstall vee deflector,
orienting apex toward gas flow. Secure with
setscrews and anti-seize compound.
Torque to 2.8 N·m (25-in-lbs). Secure hub
retaining setscrew.
k. On systems equipped with an abrasive
shield, install dust seal gaskets, with joints
180° apart.
730010
PROBE
END
FLANGE
PROBE TUBE
(NOT INCLUDED KIT)
CORRUGATED
SEAL
CELL END
FLANGE ASSEMBLY
RETAINER
SOCKET
HEAD CAP
SCREW
CALIBRATION
TUBE INSERT
CALIBRATION
TUBE PASSAGE
Figure A-12. Cell Replacement Kit