
Instruction Manual
106-300NFX Rev. 4.2
January 2002
2-22 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
HIGH CAL
GAS IN
LOW CAL
GAS IN
TEST GAS
OUT
REF AIR
OUT
INSTR
AIR
REF AIR
OUT
REF AIR
OUT
REF AIR
OUT
TEST GAS
OUT
TEST GAS
OUT
TEST GAS
OUT
PROBE 1 PROBE 2 PROBE 3 PROBE4
LINE IN
SIGNAL IN
DRAIN
35870003
Figure 2-18. MPS Gas Connections
Do not use 100% nitrogen as a low
(zero) gas. It is suggested that the low
gas be between 0.4% and 2.0% O
2
. Do
not use gases with hydrocarbon con-
centrations of more than 40 parts per
million. Failure to use proper gases
will result in erroneous readings.
3. Connect the low O
2
test gas to LOW
GAS. The test gas pressure should be
set at 138 kPa (20 psi).
4. Connect the REF AIR OUT to the ref-
erence gas fitting on the probe junction
box.
5. Connect the TEST GAS OUT to the
calibration gas fitting on the probe
junction box.
6. If the MPS is configured for multiple
probes (up to four), repeat steps 4 and
5 for each additional probe.
A check valve is required for each
probe connected to an MPS to prevent
condensation of flue gas in the cali-
bration gas lines. The check valve
must be located between the calibra-
tion fitting and the gas line.
c. Electrical Connections
Electrical connections should be made as
described in the electrical installation dia-
gram, Figure 2-19. All wiring must conform
to local and national codes. The electrical
connections will exist only between the
electronics package and the MPS to enable
automatic and semiautomatic calibration. If
more than one probe system is being used,
the additional probes and electric packages
would be wired similarly to the first probe.
NOTE
MPS power supply fuse locations and
specifications are shown in Figure
2-19.
Kommentare zu diesen Handbüchern