January 2002
3-12 Setup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
NOTE
Only set the test gas flowmeter upon
initial installation and after changing
the diffusion element. A slightly lower
test gas flow rate may indicate a
plugged diffusion element.
2 Set the test gas pressure
regulators and the flow meter
for a flow of 5 scfh at (20 psig)
138 kPa for both gases. The
reference gas should be
flowing as in normal opera-
tion.
3 Refer to paragraph 3-10d of
this section for Manual (Semi-
automatic) Calibration setup
and procedure using the IFT.
4 Test gases will be switched
on and off using the shutoff
valves.
Test Method "B" Rosemount Oxygen
Test Gas and Service Kit.
(a) Required Equipment
Do not use 100% nitrogen as a zero
gas. It is suggested that gas for the
zero be between 0.4% and 2.0% O
2
. Do
not use gases with hydrocarbon con-
centrations of more than 40 parts per
million. Failure to use proper gases
will result in erroneous readings.
1 Portable Rosemount Oxygen
Test Gas Kits (Figure 3-3),
Rosemount Part Number
6296A27G01, containing 8%
and 0.4% gases in a port-
able carrying case with regu-
lator, built-in valve, hose and
hose connecting adapter to
the calibration gas connec-
tion.
2 Extra gas bottles are available
at:
Rosemount Analytical Inc.
Box 901
Orrville, Ohio 44667
U.S.A.
Rosemount Limited
Burymead Road
Hitchin, Herts. U.K.
Rosemount Italy
VIA Guido Cavalcanti 8
20127 Milan, Italy
Rosemount Spain
Saturnino Calleja 6
28002 Madrid, Spain
Rosemount France
165 Boulevard de Vallmy
92706, Colombes, France
Rosemount Part Number
3530B07G01 for probe 0.4%
oxygen in nitrogen in dispos-
able bottle.
TEST GAS
KIT #1
(P/N 6296A27G1)
35870004
Figure 3-3. Portable Rosemount
Oxygen Test Gas Kit