Rosemount Analytical Inc. A Division of Emerson Process Management Setup and Operation 3-7
World Class 3000
Table 3-4. CALIBRATION O
2
Sub-Menu
SUB-MENU
SELECTION
SETUP SETTING
(SEE TABLE 3-4) DISPLAY DESCRIPTION
Perform
Calibration
Auto Cal in Probe
Setup is YES
Press ENTER to start Auto Calibration. MPS will start calibration probe.
Starting Automatic Calibration
High Gas _____%O
2
Time Left 0:00
Value for high O
2
test gas.
Amount of time necessary to complete
the current testing phase in min:sec.
Cell mV ______mV Cell voltage of the probe.
Low Gas _____%O
2
Time Left 0:00
Value for low O
2
test gas.
Amount of time necessary to complete
the current testing phase in min:sec.
Cell mV ______mV Cell voltage of the probe.
Resistance Check
Time Left 0:00
Resistance check in progress.
Cell _____mV _____C Cell voltage and probe temperature.
Calibration Complete
Purging 0:00
Gas lines are being purged of test gas.
Cell _____mV _____C Cell voltage and probe temperature.
Calibration Complete
Auto Cal in Probe
Setup is NO.
Press ENTER to start Manual Calibration. Manual calibration sequence will begin
when ENTER is pressed.
Switch ON high test gas. Press ENTER when
ready.
High gas ______%O
2
High O
2
test gas value.
Press ENTER when O
2
reading is stable.
Turn off high test gas and ON low test gas.
Press ENTER when ready.
Low gas ______%O
2
Low O
2
test gas value.
Press ENTER when O
2
reading is stable.
Resistance Check Resistance check in progress.
Turn off low test gas.
Press ENTER when ready.
Press ENTER when probe has returned to
process.
View
Constants
(N/A) Slope _____mV/D
Constant _____mV
Resist _____ohms
Slope for probe.
Offset for probe.
Resistance for probe.
Calibration
Status
Next Cal XD XH Time until next calibration in number of
days and number of hours.
(N/A) Slope _____
Constant _____
Resist _____
Status of the slope.
Status of the offset.
Status of the resistance.