Emerson World Class 3000 Bedienerhandbuch Seite 150

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 158
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 149
Instruction Manual
106-300NFX Rev. 4.2
January 2002
9-6 Drawings and Schematics Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
THIS DRAWING AFFECTS CENELEC CERTIFICATION.
NO REVISION ALLOWED WITHOUT CHECKING AND
APPROVAL BY THE PRODUCT ENGINEER.
DEO
DEO
05553
16-2-93
05934
4-2-94
1
2
T
IT
L
E
SIZE
SCALE
DWG NO
SHEET
SAFETY DATA SHEET FOR
CENELEC APPROVED WC3000 PROBE
1OF6
1M03226
Measurement
Control
Analytical
INTRODUCTION:
POWER DOWN PROCEDURE:
ISOLATION OF ELECTRICAL SUPPLY:
RESTORATION OF SUPPLY FOR ELECTRICAL TESTING:
REMOVAL OF JUNCTION BOX COVER:
THREADED JOINTS:
THIS SAFETY DATA SHEET APPLIES TO WC3000 INSITU OXYGEN ANALYZER
(PROBE),TYPE No. 1U05680. THIS OXYGEN ANALYZER IS CERTIFIED EExd lIB 370 (T1):
ISSeP CERTIFICATE No. 93C.103.1067.
THIS PIECE OF EQUIPMENT HAS BEEN DESIGNED AND MANUFACTURED TO OPERATE
SAFELY IN CERTAIN TYPES OF POTENTIALLY EXPLOSIVE ATMOSPHERES. IT IS ESSENTIAL
THAT THE EQUIPMENT IS NOT TAMPERED WITH OR DAMAGED IN ANY WAY WHICH
MIGHT LEAD TO A REDUCTION IN IT'S ABILITY TO OPERATE SAFELY IN SUCH POTENTIALLY
EXPLOSIVE ATMOSPHERES. FOR YOUR OWN SAFETY AND THE SAFETY OF OTHERS
PLEASE BRING ANY DAMAGE TO THE ATTENTION OF THE RESPONSIBLE AUTHORITY.
THIS PIECE OF EQUIPMENT HAS BEEN DESIGNED AND MANUFACTURED IN
ACCORDANCE WITH EUROPEAN STANDARDS EN50014 & EN50018. INSTALLATION
MAINTENANCE AND REPAIR MUST BE IN ACCORDANCE WITH THE OFFICIAL CODES
OF PRACTICE ON THE INSTALLATION AND MAINTENANCE OF ELECTRICAL APPARATUS
IN POTENTIALLY EXPLOSIVE ATMOSPHERES' FOR THE COUNTRY OF INSTALLATION
(EXAMPLE: BS5345 IN GREAT BRITAIN). ONLY APPROPRIATELY TRAINED PERSONNEL
ARE AUTHORIZED TO PERFORM ANY WORK ON THIS EQUIPMENT. SUCH PERSONNEL IN
ADDITION TO OPERATING TO THE ABOVE MENTIONED SAFETY STANDARDS, SHOULD TAKE
NOTE OF THE FOLLOWING SAFETY ISSUES.
NOTE: THROUGHOUT THIS DATA SHEET, CONTINUOUS REFERENCE WILL BE MADE TO
FIGURE 1.
(1)
THIS PIECE OF APPARATUS IS NOT ITSELF FITTED
WITH A MEANS OF ELECTRICAL ISOLATION. CONSULT YOUR LOCAL CODES OF PRACTICE
ON THE INSTALLATION AND MAINTENANCE OF ELECTRICAL APPARATUS IN POTENTIALLY
EXPLOSIVE ATMOSPHERES (BS5345 IN BRITAIN) FOR INSTRUCTION ON THE ISOLATION OF
ELECTRICAL SUPPLY TO THE APPARATUS. FURTHER MORE THERE MUST BE EFFECTIVE
MEASURES TO PREVENT THE RESTORATION OF SUPPLY TO THE APPARATUS WHILE THE
RISK OF EXPOSING UNPROTECTED LIVE CONDUCTORS TO AN EXPLOSIVE ATMOSPHERE
CONTINUES:BS5345 PART 1 1989 SECTION 18.
WHERE, FOR PURPOSES OF
ELECTRICAL TESTING, IT IS ESSENTIAL TO RESTORE THE SUPPLY BEFORE THE APPARATUS
IS REASSEMBLED, THEN THIS WORK SHOULD BE UNDER A CONTROLLED PROCEDURE
AND THE SPECIFIC LOCATION ASSESSED TO ENSURE THAT POTENTIALLY FLAMMABLE
GAS OR VAPOUR IS ABSENT:BS5345 PART 1 1989 SECTION 23.
DO NOT REMOVE THE JUNCTION BOX COVER
(ITEM 2) WHILE THE APPARATUS IS ENERGIZED. DO NOT REMOVE THE JUNCTION BOX
COVER WHEN A POTENTIALLY EXPLOSIVE ATMOSPHERE IS PRESENT. REMEMBER (ALTHOUGH
THE UNIT HAS BEEN POWERED DOWN AND THE PROBE HEATER HAS BEEN ALLOWED TO
COOL DOWN), THE TEMPERATURE ON THE INTERIOR OF THE PROBE TUBE (ITEM 1) WILL BE
SIMILAR TO THE TEMPERATURE OF THE EXHAUST GAS. IF THERE IS A POTENTIALLY
EXPLOSIVE ATMOSPHERE IN THE REGION OF THE JUNCTION BOX (ITEM 3), THIS GAS WILL,
ON THE REMOVAL OF THE JUNCTION BOX COVER BE EXPOSED TO THE HOT SURFACE OF
THE TUBE INTERIOR (ITEM 1).
(2) THE JOINT BETWEEN THE JUNCTION BOX (ITEM 3) AND THE
JUNCTION BOX LID (ITEM 2) IS A THREADED JOINT. SO ALSO IS THE JOINT BETWEEN
THE PROBE END FLANGE (ITEM 4) AND THE FLAME ARRESTOR HUB (ITEM 5). BOTH
THESE JOINTS ARE SECURED BY SET SCREWS (ITEMS 6 & 7). BEFORE REMOVING
THE JUNCTION BOX COVER (ITEM 2) OR THE FLAME ARRESTOR HUB (ITEM 5), FULLY
REMOVE THE SET-SCREWS (ITEMS6&7)FROM THEIR TAPPED HOLES. FAILURE TO
DO THIS COULD RESULT IN DAMAGE TO THE THREADS OF THE JUNCTION BOX (ITEM 3)
AND THE PROBE END FLANGE (ITEM 4) BY THE SETSCREWS BEING DRAWN OVER
THE THREADS. REMEMBER WHEN REFITTING THE JUNCTION BOX LID AND THE FLAME
ARRESTOR HUB TO LOCK IN PLACE WITH THE SETSCREWS.
THE MATERIAL OF THE JUNCTION BOX AND THE JUNCTION BOX LID IS ALUMINUM
ALLOY. SPECIAL CARE SHOULD BE TAKEN TO AVOID DAMAGE TO THE THREADS.
NOTE THAT ROSEMOUNT SUPPLIES ALLEN KEYS FOR REMOVAL & REPLACEMENT OF THE
SETSCREWS.
34990005