Instruction Manual
106-300NFX Rev. 4.2
January 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Setup and Operation 3-15
World Class 3000
(a) Press SETUP to display the
SETUP sub-menu. Select Calibra-
tion. Ensure that Auto Cal is en-
abled. Set the cursor on Auto Cal.
Press ENTER. Set Auto Cal to
YES if not already done.
(b) Press the CAL key. Select Perform
Calibration. "Press ENTER to start
Automatic Calibration" will appear
on the LCD display. Press ENTER
to start. Refer to Table 3-4. CALI-
BRATE O
2
Sub-Menu.
PROBE
(END VIEW)
IFT
TEST GAS
CHECK
VALVE
PROBE
SIGNAL CONNECTIONS
MPS-IFT
SIGNAL
CONNECTIONS
MPS
INSTRUMENT
AIR IN
NOTE: 1. THE MPS CAN BE USED WITH UP
TO FOUR PROBES. ONLY ONE PROBE
CAN BE CALIBRATED AT A TIME.
PROBE CALIBRATIONS MUST BE
SCHEDULED IN MULTIPLE PROBE
APPLICATIONS.
REFERENCE
AIR
HPS
730014
TEST GAS 1
(HIGH O )
2
TEST GAS 2
(LOW O )
2
2. HPS OPTION SHOWN.
Figure 3-5. Typical Automatic Calibration System
3