Reference Manual
00809-0100-4801, Rev FA
October 2010
Rosemount 3051S Series
5-4
Remove Interface
Assembly
The Standard Interface Assembly, Adjustment Interface Assembly, Safety
Certified Electronics Assembly (with yellow casing), or HART Diagnostics
Electronics Assembly (black casing with white label) is located in the
compartment opposite the terminal side in the PlantWeb housing. To remove
the assembly, perform the following procedure.
1. Remove the housing cover opposite the field terminal side.
2. Remove the LCD Display or Adjustment Module, if applicable. To do this,
hold in the two clips and pull outward. This will provide better access to
the two screws located on the Standard Interface Assembly, Adjustment
Interface Assembly, Safety Certified Electronics Assembly, or HART
Diagnostics Electronics Assembly.
3. Loosen the two small screws located on the assembly in the 8 o’clock
and 2 o’clock positions.
4. Pull out the assembly to expose and locate the SuperModule connector,
see Figure 5-1.
5. Grasp the SuperModule connector and pull upwards
(avoid pulling wires). Housing rotation may be required to access locking
tabs. (PlantWeb housing only)
Figure 5-1. SuperModule
connector view
Remove the
SuperModule from the
Housing
IMPORTANT
To prevent damage to the SuperModule cable, disconnect it from the
PlantWeb assembly or Junction Box terminal block before you remove the
SuperModule from the housing.
1. Loosen the housing rotation set screw with a
3
/32-inch hex wrench, then
rotate back one full turn.
2. Unscrew the housing from the SuperModule.
PlantWeb Junction Box
Housing Rotation Set
Screw (3/32-inch)
Kommentare zu diesen Handbüchern