Seitenansicht 103
Reference Manual
00809-0100-4801, Rev FA
October 2010
Rosemount 3051S Series
6-6
Inspection Visual Inspection
Not required
Special Tools
Not required
Product Repair
The 3051S is repairable by major component replacement.
All failures detected by the transmitter diagnostics or by the proof-test
must be reported. Feedback can be submitted electronically at
http://rosemount.d1asia.ph/rosemount/safety/ReportAFailure_newweb.asp.
SPECIFICATIONS The 3051S must be operated in accordance to the functional and
performance specifications provided in the 3051S Product Data Sheet
(document number 00813-0100-4801).
Failure Rate Data The FMEDA report includes failure rates and common cause Beta factor
estimates.
The report is available at
http://www2.emersonprocess.com/en-US/brands/rosemount/Safety-Products/Pages/index.aspx.
Product Life 50 years – based on worst case component wear-out mechanisms –
not based on wear-out of process wetted materials
Report any safety related product information at
http://rosemount.d1asia.ph/rosemount/safety/ReportAFailure_newweb.asp.
SPARE PARTS Additional spare parts are available in Appendix A: Specifications and
Reference Data.