Reference Manual
00809-0100-4801, Rev FA
October 2010
4-15
Rosemount 3051S Series
FIELD UPGRADES
Labeling Each housing and each SuperModule is labeled individually, so it is
imperative that the approval codes on each label match exactly during
upgrade. The label on the SuperModule reflects the replacement model code
for reordering an assembled unit. The housing labeling will only reflect the
approvals and communication protocol of the housing.
Upgrading Electronics The PlantWeb housing allows for electronics upgrades. Different electronics
assemblies provide new functionality and are easily interchanged for upgrade.
Keyed slots guide the assemblies into place, and assemblies are secured with
two provided screws. If the transmitter you are intending to upgrade does not
have a PlantWeb housing, refer to the spare parts section on page A-38 for
ordering information.
Hardware Adjustments
The D1 option is available for local hardware adjustments. This option is
available for both the PlantWeb and Junction Box housings. In order to use
zero, span, alarm and security functions, replace the existing PlantWeb
assembly with the Hardware Adjustment Interface Assembly (p/n
03151-9017-0001). Install the LCD display or hardware adjustment module to
activate the hardware adjustments.
Advanced HART Diagnostics
The DA2 option is available for Advanced HART Diagnostics. This option
requires the use of the PlantWeb housing. In order to gain full access to the
Advanced HART Diagnostic capabilities, simply add the 3051S HART
Diagnostics Electronics assembly (p/n 03151-9071-0001). Before replacing
the existing assembly with the new 3051S Diagnostics Electronics assembly,
record the transmitter configuration. Transmitter configuration data must be
reentered after adding the Advanced HART Diagnostics electronics assembly
and before putting the transmitter back into operation.
F
OUNDATION Fieldbus
F
OUNDATION fieldbus Upgrade Kits are available for PlantWeb housings. Each
kit includes an electronics assembly and terminal block. To upgrade to
F
OUNDATION fieldbus, replace the existing electronics assembly with the
F
OUNDATION fieldbus Output Electronics assembly (P/N 03151-9020-0001)
and replace the existing terminal block with the F
OUNDATION Fieldbus terminal
block (part number will vary based on the kit selected). Table 4-2 shows the
available kits.
Table 4-2. F
OUNDATION fieldbus Upgrade Kits
Kit Part Number
Standard F
OUNDATION fieldbus Upgrade Kit 03151-9021-0021
Transient Protection FOUNDATION fieldbus Upgrade Kit 03151-9021-0022
FISCO F
OUNDATION fieldbus Upgrade Kit 03151-9021-0023
Refer to “Disassembly Procedures” on page 5-3 for information on assembly.