18
Reference Manual
00809-0100-4803, Rev DA
Section 3: Configuration
April 2013
Configuration
Note
The following example will show a flow configuration for an application with Database Gas Air as
the process fluid and a 405C Conditioning Orifice Plate as the primary element. The procedure
to configure an application with other fluids and other primary elements will be similar to this
example. Natural gases, custom liquids, and custom gases require additional steps during the
configuration. See “Other fluid configurations” on page 26 for more information.
1. Engineering Assistant may open to the Preferences tab. Using the tabs at the top of the
screen, navigate to the Fluid Selection tab.
2. Expand the Gas category (click on the + icon).
3. Expand the Database Gas category.
4. Select the appropriate fluid (Air for this example) from the list of database fluids.
Figure 3-7. Fluid selection tab - database gas air
5. Enter the Nominal Operating Pressure, press the Enter or Tab key.
Note
The Nominal Operating Pressure must be entered in absolute pressure units.
6. Enter the Nominal Operating Temperature, press the Enter or Tab key. Engineering
Assistant will automatically fill in suggested operating ranges, as shown in Figure 3-7.
These values may be edited as needed by the user.