Emerson 3051S Spezifikationen Seite 117

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 188
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 116
27
Reference Manual
00809-0100-4803, Rev DA
Section 4: Operation and Maintenance
April 2013
Operation and Maintenance
4.6.4 LCD display
Transmitters ordered with the LCD display will be shipped with the display installed. Installing
the display on an existing 3051S MultiVariable Transmitter requires the LCD display kit (part
number 03151-9193-0001 for aluminum housing and 03151-9193-0004 for stainless steel
housing).
Use the following procedure and Figure 4-13 to install the LCD display:
1. If the transmitter is installed in a loop, then secure the loop and disconnect power.
2. Remove the transmitter cover on the feature board side (opposite the field terminals
side). Do not remove the instrument covers in explosive environments when the circuit
is live.
3. Engage the four-pin connector into the feature board and snap the LCD display into
place.
4. Install the display cover and tighten to insure metal to metal contact in order to meet
flameproof/explosion-proof requirements.
Figure 4-13. Optional LCD display
4.6.5 Flange and drain vent
The 3051S MultiVariable Transmitter is attached to the process connection flange by four bolts
and two alignment cap screws.
1. Remove the two alignment cap screws.
Figure 4-14. Alignment cap screws
2. Remove the four bolts and separate the transmitter from the process connection, but
leave the process connection flange in place and ready for re-installation.
LCD Display
DisplayCover
Feature Board
Alignment Cap Screw