Emerson MICRO MOTION 2700 Installationshandbuch Seite 167

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 264
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 166
Transmitter Configuration and Use: Series 1000 and 2000 Transmitters 157
Enhanced Density Model 2500 CIOModel 1500 ANCalibration
Appendix C
Calibrating the Transmitter
C.1 Overview
This section describes transmitter calibration procedures. Using these procedures, you will be able to:
Calibrate for density
Calibrate for temperature
Note: All ProLink II procedures provided in this section assume that your computer is already
connected to the transmitter and you have established communication. All ProLink II procedures also
assume that you are complying with all applicable safety requirements. See Chapter 3 for more
information.
Note: All HART Communicator key sequences in this section assume that you are starting from the
“Online” menu. See Chapter 4 for more information.
C.2 About calibration
The flowmeter measures process variables based on fixed points of reference. Calibration adjusts
those points of reference.
The transmitter is factory calibrated and does not normally need to be calibrated in the field. Calibrate
the transmitter only if you must do so to meet regulatory requirements.
Note: Micro Motion recommends using meter factors, rather than calibration, to prove the meter
against a regulatory standard or to correct measurement error. Contact Micro Motion before
calibrating your flowmeter. For information on meter factors, see Section 8.13.
C.3 Density calibration
Density calibration includes the following calibration points:
All sensors:
- D1 calibration (low-density)
- D2 calibration (high-density)
T-Series sensors only:
- D3 calibration (optional)
- D4 calibration (optional)
For T-Series sensors, the optional D3 and D4 calibrations could improve the accuracy of the density
measurement. If you choose to perform the D3 and D4 calibration:
Do not perform the D1 or D2 calibration.
Perform D3 calibration if you have one calibrated fluid.
Perform both D3 and D4 calibrations if you have two calibrated fluids (other than air and
water).
Model 1500 AN
Model 1700 AN
Model 1700 IS
Model 2500 CIO
Model 2700 AN
Model 2700 IS
Model 2700 CIO
Seitenansicht 166
1 2 ... 162 163 164 165 166 167 168 169 170 171 172 ... 263 264

Kommentare zu diesen Handbüchern

Keine Kommentare