Installation and Use Manual 83
Configuring the Discrete Batch Application
Index Discrete Batch Index Discrete Batch Index Discrete Batch Index Discrete Batch
If the Inhibit Totalizer function is mapped to a discrete input:
The totalizer will not increment while the discrete input is active.
The totalizer will resume incrementing automatically when the discrete input becomes
inactive.
If the discrete input is inactive, the totalizer is active, whether or not a batch is running. Even if
a batch is not running, the totals will be incremented if flow is detected.
If the batch totalizer is inhibited while a batch is running:
•And the
No-flow timeout batch control option (see Section B.3.2) is set to 0, no timeout alarm
will be posted.
•And the
No-flow timeout batch control option is set to a non-zero value, a timeout alarm will
be posted if the timeout period expires before batch totalizing resumes.
B.4 Running a batch
The ProLink II
Run Batcher window, shown in Figure B-4, can be used for batch operation. It
provides the same functionality as the batch process screen on the transmitter display.
Figure B-4 Run Batcher window