
Installation and Use Manual 53
Transmitter Configuration, Characterization, and Calibration
Meter Verification Transmitter TerminalsData LoggerTransmitter Configuration
Figure 5-6 D3 or D3 and D4 density calibration
Fill sensor with D3 fluid
Calibration
Density cal - Point 3
Do Cal
Enter density of D3 fluid
Fill sensor with D4 fluid
Calibration in Progress
light turns red
ProLink
Calibration in Progress
light turns green
Calibration
Density cal - Point 4
Do Cal
Enter density of D4 fluid
Calibration in Progress
light turns red
ProLink
Calibration in Progress
light turns green
Close shutoff valve
downstream from sensor
Done
Close Close
Kommentare zu diesen Handbüchern