42 ProLink
®
II Software for Micro Motion
®
Transmitters
Initial Transmitter Startup Procedures
Figure 4-4 Zeroing
4. Change the
Zero Time if necessary. You will need to click Apply to send the new value to the
transmitter before performing the actual zeroing procedure.
5. If the Convergence Limit parameter is displayed, type a new convergence limit in the
Converg. Limit box or accept the default value.
6. Click
Perform Auto Zero. The meter will begin zeroing. The Calibration in Progress status
light will turn red.
7. Wait until the zero time has expired. At the end of this time:
If the
Calibration Failure status light turns red, the zero procedure failed.
- See your transmitter manual for troubleshooting procedures.
- If desired, use the buttons in the dialog box to restore the previous zero value or the
zero value established during factory calibration.
Note: These two functions are not available on all transmitters.
If the
Calibration in Progress status light returns to green and the Calibration Failure
status light does not turn red, the zero procedure succeeded.
8. Click
Close.