STARTUP and OPERATION
zone may show than the AC/Heater unit fan may have failed or a possible defective
thermocouple.
*PDD Temp: For systems with NOx a cylindrical NOx detector assy is located in the
detector section. Internal to the detector assy a small peltier device is operating and
must operate at 0 degrees C. The temperature should never deviate +/- .05 degrees C
from the setpoint of zero or the NOx readings may drift. Integral will typically run
between 40 to 70%.
*PMT Temp: This temperature is for the detector assy heater core. Setting is set to 52
degrees C. Temperature should not drift more than 0.2 degrees C or NOx drift may
occur. Excessive temperature variation may be caused by either poor zone temperature
control or a faulty heater.
*Conv Temp: This temperature is for the NOx converter assembly. Temperature setting
is 330 degrees C. Temp should not vary more than 1 degrees C. or NOx
measurements will drift. A faulty heater will cause temp variations.
Note that when the enclosure door is opened that all of the above temperature setting
may be affected and will take a short about of time to react and control to the desired
temperatures.
Figure 4.16 - View Data Logs
Table 4.10 - Average Period Selection
Note: The page header
was scrolled out of view
to show all the selection
options, but it can be
seen in Fi
ure 4.17
If Most Recent is
selected, the month day
and hour do not need to
be selected.
Select the ending hour
to view (applicable only
to 1 minute averages)
Select 1 min., 15 min.,
1 hour or 24 hour
averages.
Average Period Time Range
Displayed
1 Minute 1 Hour
15 Minutes 1 Day
1 Hour 3 Days
24 Hours 3 Months
Rosemount Analytical µCEM Continuous Analyzer Transmitter 4-2