INSTALLATION
J VSYNC BLACK CENTER
K GND BLACK SHIELD
L DATA DCC DATA
M CLK DCC CLK
N KBDATA KEYBOARD DATA
P KBCLK KEYBOARD CLOCK
R GND GROUND
S VCC VCC, +5VDC
R GND GROUND
S VCC VCC, +5VDC
T MSDATA MOUSE DATA
U MSCLK MOUSE CLOCK
Table 3-5. CPU I/O Terminal Assignments
3.3.4.7 SSU Power Connector, T/S units Only (J8) – T/S enclosure can be located away from
the Analysis enclosure. This cable serves as the connection and is by Rosemount.
The SSU Power connector has a shell size of 8, 3 contacts. Each pin will accept a wire size of
24, 22, or 20 AWG. Connector and 6’ pigtail by Rosemount.
SIGNAL NAME DEFINITION PIN
SSUCtrl
SSU Control line
A
Vbb_rtn
+24V Return
B
Gnd
GND
C
Table 3-6. SSU Power Connection Terminal Assignments
3.3.4.8 AC Power Connector (J1) – Customer 120VAC Power Connection
The AC Power Interface connector has a shell size of 12, 3 contacts. Each pin will accept a
wire size of 16 AWG. Connector and 6’ pigtail by Rosemount.
SIGNAL NAME DEFINITION PIN
L1 A
L2
85-264 VAC, 47-440 Hz
C
GND AC Ground B
Table 3-7. AC Power Connection Terminal Assignments
Connect AC power through a 20A circuit breaker that is to be located close to the
µCEM. The circuit breaker will provide over current protection as well as a means of
disconnecting the power.
Maximum power requirements will be 1000 watts, with most applications requiring less
than this amount
Rosemount Analytical µCEM Continuous Analyzer Transmitter 3–2
Kommentare zu diesen Handbüchern