STARTUP and OPERATION
Table 4.3 - [General] section
Calibration Setting Description
Stream1Time Stream 1 processing time in minutes when auto switching
Stream2Time Stream 2 processing time in minutes when auto switching
TransitionTime Time to keep the B flag after the switch have occurred, in
seconds
Stream1Name Stream 1 Name to be shown on Pocket PC and Web pages
Stream2Name Stream 2 Name to be shown on Pocket PC and Web pages
CalibrationCurrentLoopOutputs Defines the behavior of Current Loops during Calibrations
1 - Hold the Last Good Value,
2 - Use the User-Specified Value
3 - Follow the Gases as is
CalibrationCurrentLoopOutputsUserValue Value in milliamps. Used when the previous parameter is
set to 2
ByPassCurrentLoopOutputs Defines the behavior of Current Loops during By-Pass
1 - Hold the Last Good Value
2 - Use the User-Specified Value
ByPassCurrentLoopOutputsUserValue Value in milliamps. Used when the previous parameter is
set to 2
AutoCalForcesSwitch Defines what to do, when the scheduled Auto-Calibration
time comes, but the system happens to process another
stream
1 - force a switch to the stream and run the Calibration
2 - wait until the stream is switching occures by itself and
run the Calibration
DigitalOutputsLogic Defines how to control Digital Outputs
1- O2 Limit, CO Limit, NOX Limit Logic
2- Valid, In Calibration, In Maintenance
Rosemount Analytical µCEM Continuous Analyzer Transmitter 4-5