26 Micro Motion
®
ELITE
®
Flow and Density Meters
Hazardous area classifications continued
Model CMF400 with core processor
Use the above graph to determine the temperature class for a given fluid and ambient temperature. The maximum surface
temperature for dust is as follows: T5:T 95°C, T4:T 130°C, T3:T 195°C, T2 to T1:T 234°C.
Ambient temperature range Ta –40 °C to +60 °C
High-temperature models CMF200A, CMF200B, CMF300A, CMF300B, CMF400A, or CMF400B with
core processor or Model 1700/2700 transmitter
Use the above graph to determine the temperature class for a given fluid and ambient temperature. The maximum surface
temperature for dust is as follows: T5:T 95 °C, T4:T 130 °C, T3:T 195 °C, T2: T 290 °C, T1:T 363 °C. The minimum ambient and
process fluid temperature allowed for dust is –40 °C.
Since the electronics are mounted approx. 1 meter away from the sensor by means of a flexible stainless steel hose, the use of the
sensor at an ambient temperature higher than +55 °C is possible, provided that the ambient temperature does not exceed the
maximum temperature of the medium taking into account the temperature classification and the maximum operating temperature of
the sensor.
Ambient temperature range Ta –50°C to +55°C
Sensor fluid temp (°C)
Max ambient temp (°C)
Sensor fluid temp (°C)
Max ambient temp (°C)