8 Micro Motion
®
Model 2400S Transmitters
Environmental limits
Environmental effects
Hazardous area classifications
Ambient temperature limits Operating and storage: –40 to +140 °F (–40 to +60 °C)
Below –4 °F (–20 °C), LCD responsiveness decreases and LCD may become difficult to
read. Above 131 °F (55 °C), some darkening of the LCD panel
might occur.
ATEX requires limiting ambient temperature to below 131 °F (55 °C).
Humidity limits 5 to 95% relative humidity, non-condensing at 140 °F (60 °C)
Vibration limits Meets IEC68.2.6, endurance sweep, 5 to 2000 Hz, 50 sweep cycles at 1.0 g
All models
EMI effects Complies with EMC Directive 2004/108/EC per EN 61326 Industrial
Complies with NAMUR NE-21 (22.08.2007)
Model 2400S Analog
Ambient temperature effect On mA output: ±0.005% of span per °C
CSA C-US Class I Division 2 Groups A, B, C, D
Class II Division 2 Groups F and G
ATEX
Analog or
PROFIBUS-DP
II 3 G Ex nAC II T5 (Zone 2)
II 3 D IP66/IP67 T70°C
DeviceNet II 3 G Ex nA II T5 (Zone 2)
II 3 D IP66/IP67 T70°C
IECEx
All models Ex nAC II T5
INMETRO
Analog
BR-Ex nAC IIC T5
–40 °C Ta +60 °C