Micro Motion
®
Model 2400S Transmitters 3
Physical specifications
Power supply – Analog and PROFIBUS-DP
Electrical connections
Housing NEMA 4X (IP66/67) polyurethane-painted cast aluminum or 316L stainless steel
Weight Transmitter is mounted integrally with sensor. For weight of flowmeter, see product data
sheet for sensor.
Mounting Model 2400S transmitters are mounted integrally with sensor. The transmitter can be rotated
on the sensor up to 360° in 45° increments.
Self-switching AC/DC input, automatically recognizes supply voltage.
Complies with low voltage directive 2006/95/EC per EN 61010-1 (IEC 61010-1).
Installation (Overvoltage) Category II, Pollution Degree 2.
AC • 85–265 VAC
• 50/60 Hz
• 4 watts typical, 7 watts maximum
DC • 18–100 VDC
• 4 watts typical, 7 watts maximum
Fuse • IEC 127-1.25 fuse, slow blow
Model 2400S Analog
Input and output connections Two pairs of wiring terminals for transmitter inputs/outputs. Screw terminals accept solid or
stranded conductors, 26 to 14 AWG (0.14 to 2.5 mm
2
).
Power connections One pair of wiring terminals accepts AC or DC power. One internal ground lug for power
supply ground wiring.
Screw terminals accept solid or stranded conductors, 26 to 14 AWG (0.14 to 2.5 mm
2
).
Digital communications
maintenance connections
Two clips for temporary connection to the service port.
Two clips for temporary connection to HART/Bell 202 terminals.
Model 2400S PROFIBUS-DP
PROFIBUS-DP segment One pair of wiring terminals for connection to PROFIBUS-DP segment. Connection type:
• Screw terminals accept solid or stranded conductors, 26 to 14 AWG (0.14 to 2.5 mm
2
).
• Five-pin PROFIBUS-DP M12 (Eurofast) female connector (optional).
Power connections One pair of wiring terminals accepts AC or DC power. One internal ground lug for power
supply ground wiring.
Screw terminals accept solid or stranded conductors, 26 to 14 AWG (0.14 to 2.5 mm
2
).
Digital communications
maintenance connections
Two clips for temporary connection to the service port.
Model 2400S DeviceNet
DeviceNet segment One pre-installed male 5-pin Eurofast connector for I/O and power
supply wiring.
Digital communications
maintenance connections
Two clips for temporary connection to the service port.