13/12/93 1
DEDQ5937
7-4-94 2
(6) EXlSTlNG WESTlNGHOUSE/ROSEMOUNT SUPPLIED CABLE:
ON EARLIER INSTALLATIONS WESTINGHOUSE/ROSEMOUNT SUPPLIED A CABLE
BEIWEEN THE DIGITAL ELECTRONICS AND PROBE (MODEL 218).
THE GLANDS ON THIS CABLE - ALTHOUGH HAWKE GLANDS CERTIFIED Exd IIC
WERE NOT OF THE HAWKE BARRIE:R (STOPPER) GLAND, MPE 1cG653 VARIETY.
THESE GLANDS MUST NOT BE USED. ROSEMOUNT CAN SUPPLY A BARRIER GLAND
KIT TO REPLACE SUCH GLANDS. THE KIT (PART No. lUO3066GO7), COMPRISES
OF TWO BARRIER GLANDS COMPLETE WITH PUl-fY, CRIMPS AND AN INSTRUCTION
SHEET. PLEASE NOTE THAT YOU SHOULD USE ONLY PUTIY WHICH IS PLASTIC
& CAPABLE OF BEING MIXED. YOU SHOULD NOT USE CRYSTALISED OR HARD
PUI-I-Y.
(7) CABLE ANCHORAGE
THIS APPARATUS DOES NOT INCLUDE ANY SPEClflC MtiNS FOR CABLE ANCHORAGE.
THE CABLE GLANDS CHOSEN FOT THE MAINS, PROBE, AND SIGNAL CABLES MUST
BE OF AMPE WHICH SUPPLY CABLE ANCHORAGE. THE BARRIER (STOPPER) GLAND
SUPPLIED BY ROSEMOUNT ASi.PART OF THE PROBE CABLE ASSEMBLY, WILL GIVE
SUCH CABLE ANCHORAGE.
(8) UNUSED CABLE ENTRY PORTS:
ALL CABLE ENTRY PORTS WHICH ARE UNUSED
ARE TO BE CLOSED OFF WITH THREADED PLUGS CERTlflED Exd IIC. THERE SHOULD
BE A MINIMUM OF 5 THREADS ENGAGED AND THE THREADS SEALED WITH
THREADLOCK (LOCKTlTE 271 OR IIQUILAVENT).
(9) NAMEPLATE (LABEL):
ENSURE THAT NAMEPLATE IS AT ALL TIMES VISIBLE.
AT INSTALLATION. ALLOWANCE MUST BE MADE FOR THIS.
(10) EARTHING OF THE APPARATUS:
THE APPARATUS HAS BEEN FITTED WITH BOTH
EXTERNAL AND INTERNAL EARTHING POINTS. ADEQUATE EARTH
CONNECTIONS
SHOULD BE MADE TO BOTH THESE POINTS.
(11) SHEET METALWORK PANELS.
THIS APPARATUS CONTAINS A NUMBER OF SHEET METALWORK PANELS FOR MOUNTING
THE PRINTER CIRCUIT BOARDS. BECAUSE THESE PANELS ARE IN CLOSE PROXIMIlY
TO THE WALLS OF THE FLAMEPROOF ENCLOSURE, IT WAS NECESSARY TO
PERFORATE THEM WITH HOLES h;ND SLOTS. THESE HOLES AND SLOTS PREVENT
PRESSURE PILING BETWEEN THE PANELS AND THE WALLS & LID OF THE
ENCLOSURE. DO NOT DO ANYTHING WHICH MIGHT OBSCURE THESE SLOTS & HOLES.
(12) LlFl-lNG AND CARRYING.
THE In3000 IS A HEAVY PIECE OF APPARATUS. LIFTING AND CARRYING PROCEDURES
SHOULD TAKE ACCOUNT OF THIS WEIGHT.
(13) CONNECTION TO HART OPTION.
HART IS A COMMUNICATIONS PROTOCOL WHICH ALLOWS REMOTE COMMUNICATION
WITH THE lFr3000 VIA THE 4-20mA OUTPUT. BECAUSE, ON THE IFT3000, THE
HART OPTION IS NOT PROTECTED BY ENERGY LIMITING BARRIERS. IT MUST NOT
BE INTERFACED FROM WITHIN THE HAZARDOUS AREA. THE 4-20mA CABLES SHOULD
BE ROUTED OUTSIDE THE HAZARDOUS AREA, AND THE CONNECTION MADE OUT
THERE. NOTE THIS IS THE CASE EVEN WHEN USING THE INTRINSICALLY SAFE
VERSION OF THE THE HANDHELD COMMUNICATOR. THE LIMITATION (ie, NO ENERGY
LIMITING BARRIER) LIES IN THE IFT3000 NOT IN THE HANDHELD UNIT.