Model 1054B DCDual CellConductivity/ResistivityMicroprocessor AnalyzerInstruction ManualPN 5101054BDCJanuary 2000FREEEXTENDED WARRANTY! (See offer at
5MODEL 1054B DC SECTION 2.0INSTALLATIONSECTION 2.0INSTALLATION2.1 GENERAL. This analyzer's enclosure is suitablefor outdoor use. However, it shou
6MODEL 1054B DC SECTION 2.0INSTALLATION2.4.1 Power Input Wiring. The Model 1054B can beconfigured for either 115 VAC or 230 VAC power.Connect AC power
7MODEL 1054B DC SECTION 2.0INSTALLATIONFIGURE 2-1. Panel Mounting CutoutWHEN INCH AND METRIC DIMSARE GIVENMILLIMETERINCHDWG. NO. REV.41054B01 B
8MODEL 1054B DC SECTION 2.0INSTALLATIONFIGURE 2-2. Panel Mounting Tab InstallationDWG. NO. REV.41054A26 A
9MODEL 1054B DC SECTION 2.0INSTALLATIONFIGURE 2-3. Pipe Mounting InstallationWHEN INCH AND METRIC DIMSARE GIVENMILLIMETERINCHDWG. NO. REV.41054B02 C
10FIGURE 2-4. Electrical WiringMODEL 1054B DC SECTION 2.0INSTALLATIONDWG. NO. REV.41054B46 A
11FIGURE 2-5. Sensor WiringMODEL 1054B DC SECTION 2.0INSTALLATIONNOTE: TERMINALS IN JUNCTION BOX ARENOT NUMBERED.✳✳OLD MODEL 140 SERIES SENSORS MAYNOT
12Configuration is all accomplished through a series ofmenus located within the set mode menu. To accessthese set mode menus the ACCESS keypad ispres
13MODEL 1054B DC SECTION 3.0DESCRIPTION OF CONTROLSFIGURE 3-4. LCD DisplayRELAY 1ACTIVATEDRELAY 3ACTIVATED(MINUS SIGN)RELAY 2ACTIVATEDCELL SELEC-TION:
14TABLE 3-1. Key DescriptionDisplays - current output (mA or % full scale).Set Function (w/SELECT) - Simulates currentoutput.Displays - low current se
WARNINGELECTRICAL SHOCK HAZARDMaking cable connections to and servicing thisinstrument require access to shock hazard level volt-ages which can cause
15MODEL 1054B DC SECTION 3.0DESCRIPTION OF CONTROLSMNEMONIC DESCRIPTION1°C Temperature °C Cell 12°C Temperature °C Cell 21°F Temperature °F Cell 12°F
16MODEL 1054B DC SECTION 3.0DESCRIPTION OF CONTROLS1Cin2Cin1CEL2CEL1FCt2FCt1tYP2tYPAL1AL2AL31t-C2t-C1OUt2OUtUErdFtdtSCOdCYCSEt1t2tOFFHiLoSP3H-LHY5dond
17MODEL 1054B DC SECTION 4.0CONFIGURATIONSECTION 4.0CONFIGURATION4.1 GENERAL. This section details all of the itemsavailable in the Set Mode to config
18MODEL 1054B DC SECTION 4.0CONFIGURATIONTABLE 4-1.CONFIGURATION WORKSHEETUse this worksheet to assist in the configuration of the analyzer. Date: ___
19D. Alarm Logic. Mnemonic H-L. Select this item forhigh or low alarm logic. High logic activates the alarmwhen the reading is greater than the set po
20MODEL 1054B DC SECTION 4.0CONFIGURATION5. If OFF is selected, display will show oFF toacknowledge. Press ENTER key to return to AL1 orAL2, concludi
21MODEL 1054B DC SECTION 4.0CONFIGURATION9. SCROLL(é) to the desired item and ENTER it intomemory. Display will return to H-L. If changes torelay acti
224.5.1 Current Output Configuration 1OUt / 2OUt. Refer toFigure 4-5.1. Enter Set Mode by pressing the ACCESS key twice.2. SCROLL (é) until 1OUt/2OUt
232. Press the ALARM 1 or ALARM 2 key. SP1 or SP2will show briefly, followed by the Alarm 1 or Alarm2 Setpoint.NOTE:If the alarm is set to OFF or FLt
24MODEL 1054B DC SECTION 4.0CONFIGURATION4.8 OUTPUT SCALE EXPANSION. The output is con-trolled as user configured. The output zero and fullscale value
254.9 SIMULATE CURRENT OUTPUTS. The outputscan be simulated to check the operation of devicessuch as valves, pumps, or recorders. The outputs canbe si
264.10 DISPLAY CYCLE. Display Mnemonic CYC. Thisfeature allows the display to alternate between cellone and cell two PV readings. Refer to Figure 4-10
27MODEL 1054B DC SECTION 5.0START-UP AND CALIBRATIONSECTION 5.0START-UP AND CALIBRATION5.1 START-UP AND CALIBRATION. Calibration andoperation of the M
28NOTEAfter initial installation or any cleaning pro-cedure, the conductivity cell will require Ito 2 days after insertion into the ultrapurewater str
296.1 KEYBOARD SECURITY. Display Mnemonic COd.Select this feature to display the user defined securitycode. Any three digit number may be used for thi
307.1 THEORY OF OPERATION. This section is a gen-eral description of how the analyzer operates. Thissection is for those users who desire a greater un
31MODEL 1054B DC SECTION 8.0DIAGNOSTICS AND TROUBLESHOOTINGSECTION 8.0DIAGNOSTICS AND TROUBLESHOOTINGDisplay DescriptionEEP EEPROM write error (bad EE
32MODEL 1054B DC SECTION 8.0DIAGNOSTICS AND TROUBLESHOOTING8.2 TROUBLESHOOTING. The Model 1054B analyzeris designed with the state of the art micropro
33MODEL 1054B DC SECTION 8.0DIAGNOSTICS AND TROUBLESHOOTING4. If the conductivity cell electrodes have becomeextremely dirty or fouled, chemical clean
34MODEL 1054B DC SECTION 8.0DIAGNOSTICS AND TROUBLESHOOTINGTABLE 8-3. Troubleshooting GuideSYMPTOM PROBLEM ACTIONAnalyzer value not the same 1. Refere
MODEL 1054B DC TABLE OF CONTENTSMODEL 1054B DC CONDUCTIVITYMICROPROCESSOR ANALYZERTABLE OF CONTENTSSection Title Page11..00 DDEESSCCRRIIPPTTIIOONN AAN
35FIGURE 8-1. Simulate Conductivity InputMODEL 1054B DC SECTION 8.0DIAGNOSTICS AND TROUBLESHOOTING8.2.5 Instrument Electronic Check. This procedurewil
36MODEL 1054B DC SECTION 9.0RETURN OF MATERIALSECTION 9.0RETURN OF MATERIAL9.1 GENERAL.To expedite the repair and return of instruments, prop-er commu
FROM: RETURN BILL TO:_____________________________ _____________________________ __________________________________________________________ __________
U.S. Field ServiceField Watch Response Center1-800-654-7768InternationalARGENTINAEmerson ArgentinaBuenos Aires54 1 780311,795-9338AUSTRALIAFisher-Rose
FOLD ALONG DOTTED LINES
DOUBLE WARRANTY REGISTRATIONDate of Purchase (of instrument) __________________________ Your Name _______________________________________Company Name
WARRANTYGoods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in work-manship and material under norm
Rosemount Analytical Inc.Uniloc Division2400 Barranca ParkwayIrvine, CA 92606 USATel: (949) 863-1181http://www.RAuniloc.com© Rosemount Analytical Inc
MODEL 1054B DC TABLE OF CONTENTSTABLE OF CONTENTS CONT'D.LIST OF FIGURESFigure No. Title Page2-1 Panel Mounting Cutout...
1MODEL 1054B DC SECTION 1.0DESCRIPTION AND SPECIFICATIONSSECTION 1.0DESCRIPTION AND SPECIFICATIONS1.1 FEATURES AND APPLICATIONSThe Model 1054B Micropr
MODEL 1054B DC SECTION 1.0DESCRIPTION AND SPECIFICATIONS21.2 PHYSICAL SPECIFICATIONS -GENERALEnclosure: Black ABS, with interior conductive coating,NE
3MODEL 1054B DC SECTION 1.0DESCRIPTION AND SPECIFICATIONS1.3 ANALYZER SPECIFICATIONS @ 25°CMeasurement Range: 0-20,000 µS/cm or 0-50.00 megohms-cmOutp
4MODEL 1054B DC SECTION 1.0DESCRIPTION AND SPECIFICATIONS1054B DC REPLACEMENT PARTS AND ACCESSORIESPN ACCESSORIES2001492 Tag, Stainless Steel, Specify