Emerson ROSEMOUNT 5300 Installationsanleitung Seite 1

Stöbern Sie online oder laden Sie Installationsanleitung nach Messen, Testen & Steuern Emerson ROSEMOUNT 5300 herunter. Emerson ROSEMOUNT 5300 Installation guide Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken

Inhaltsverzeichnis

Seite 1 - Guided Wave Radar

Quick Installation Guide00825-0100-4530, Rev ECMarch 2013Superior Performance Guided Wave Radar Level and Interface Transmitter

Seite 2 - About this guide

March 201310Quick Installation GuideStep 3 continued...HART communicationThe Rosemount 5300 Series transmitter operates with a power supply ranging fr

Seite 3 - Contents

Quick Installation Guide11March 2013Step 3 continued...Type n approvals : non-sparking / energy limited power supplyTransient protection terminal bloc

Seite 4 - Switch HART revision mode

March 201312Quick Installation GuideStep 3 continued...Load limitationsThe Field Communicator requires a minimum load resistance of 250  within the l

Seite 5 - Tri-Clamp tank connection

Quick Installation Guide13March 2013Step 3 continued...FOUNDATION fieldbusThe Rosemount 5300 Series transmitter, FOUNDATION fieldbus version, operates

Seite 6 - Bracket mounting, on pipe

March 201314Quick Installation GuideStep 3 continued...Type n approvals : non-sparking / energy limited power supplyRS-485 with Modbus communicationTh

Seite 7 - Remote housing

Quick Installation Guide15March 2013Step 4: ConfigureNOTE:If the transmitter is pre-configured at the factory, it is only necessary to proceed with th

Seite 8 - Step 3: Connect the wiring

March 201316Quick Installation GuideConfiguration using the Rosemount Radar Master software1. Start Rosemount Radar Master (Programs > Rosemount &g

Seite 9 - To connect the transmitter

Quick Installation Guide17March 2013are normally configured at the factory, but can be changed if, for example, the probe is cut in the field, or if d

Seite 10 - HART communication

March 201318Quick Installation GuideClick Next and the following window appears (Tank Environment).8. In the Environment window, select Measurement Mo

Seite 11 - Step 3 continued

Quick Installation Guide19March 2013the available pre-defined tank options, or if high volume accuracy is desired.Choose None if volume calculation is

Seite 12 - Load limitations

March 20132Quick Installation GuideAbout this guide This installation guide provides basic guidelines for Rosemount 5300 Series transmitters. It does

Seite 13 - FOUNDATION fieldbus

March 201320Quick Installation Guide12. The Basic Configuration with the Radar Master Wizard is now finished. Proceed with the Device Specific Setup t

Seite 14 - termination

Quick Installation Guide21March 2013Safety Instrumented Systems (4-20 mA only)The following section is applicable to the Rosemount 5300 Prior-Use opti

Seite 15 - Step 4: Configure

March 201322Quick Installation GuideDampingUser adjusted damping affects the transmitter’s ability to respond to process changes. Therefore, the dampi

Seite 16 - March 2013

Quick Installation Guide23March 2013Site acceptanceAfter the installation and configuration, proper operation of the transmitter should be verified. A

Seite 17

March 201324Quick Installation Guide6. Enable write protection.7. Restore the loop to full operation.8. Remove the bypass from the safety logic solver

Seite 18

Quick Installation Guide25March 2013Product certificationsEU conformityThe EC declaration of conformity can be found on page 33. The most recent revis

Seite 19

March 201326Quick Installation GuideI5, IE1 Intrinsically safe and FISCO model:Intrinsically Safe for Class I, II, III, Div. 1, Groups A, B, C, D, E,

Seite 20

Quick Installation Guide27March 2013European certificationsATEX approvalsSpecial conditions for safe use (X)The intrinsically safe circuits do not wit

Seite 21 - Configuration

March 201328Quick Installation GuideBrazilian certificationsINMETRO approvalsSpecial conditions for safe use (X)Letter X in the number of the certific

Seite 22 - Write protection

Quick Installation Guide29March 2013I21 Intrinsically safe:4-20 mA/HART model:Ex ia IIC T4 GaEx ta IIIC T79 °C-50 °C < Ta < +70 °CUi: 30 V, Ii:

Seite 23 - Operation and maintenance

Quick Installation Guide3March 2013Any substitution of non-authorized parts or repair, other than exchanging the complete transmitter head or probe as

Seite 24 - References

March 201330Quick Installation GuideIC1 FOUNDATION fieldbus FISCO model:Ex ia IIC T4 (-50 °C < Ta< +60 °C) DIP A20 TA69 °CUi = 17.5 V, Ii = 380

Seite 25 - Product certifications

Quick Installation Guide31March 2013IECEx certificationsIECEx approvalsSpecial conditions for safe use (X)The intrinsically safe circuits do not withs

Seite 26 - fieldbus or FISCO option

March 201332Quick Installation GuideOther certificationsOverfill protectionCert. no: Z-65.16-476U1 TÜV-tested and approved by DIBt for overfill protec

Seite 27 - ATEX approvals

Quick Installation Guide33March 2013EC Declaration of ConformityFigure 1. EC Declaration of Conformity - page 1

Seite 28 - INMETRO approvals

March 201334Quick Installation GuideFigure 2. EC Declaration of Conformity - page 2

Seite 29 - Protection

Quick Installation Guide35March 2013Figure 3. EC Declaration of Conformity - page 3

Seite 30 - Japanese certifications

Emerson Process ManagementRosemount Measurement 8200 Market Boulevard Chanhassen, MN 55317 USA Tel (USA) 1 800 999 9307Tel (International) +1 952 906

Seite 31 - IECEx approvals

March 20134Quick Installation GuideStep 1: Confirm system readinessConfirm HART revision capability If using HART-based control or asset management s

Seite 32 - Overfill protection

Quick Installation Guide5March 2013Step 2: Mount the transmitter head/probeTank connection with flange1. Place a gasket on top of the tank flange. 2

Seite 33 - EC Declaration of Conformity

March 20136Quick Installation GuideStep 2 continued...Bracket mounting, on wall1. Mount the bracket directly to the wall with screws suitable for th

Seite 34

Quick Installation Guide7March 2013Step 2 continued...Remote housing1. Remove the transmitter head from the probe by unscrewing the M50 nut.For safe

Seite 35

March 20138Quick Installation GuideStep 3: Connect the wiringWiring and power supply requirements can be dependent upon the approval certification. As

Seite 36 - 00825-0100-4530

Quick Installation Guide9March 2013To connect the transmitter1. Verify that the power supply is disconnected.2. Remove the terminal block cover (see p